Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa uR1000: (1 year) CAL VER /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 24 NUMBER OF LINES: 134 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DC VOLTAGE} 1.007 OPBR Is the UUT the 6-DOT printing model? 1.008 JMPT 1.011 # Pen type inputs 1.009 DISP Connect 5520A to the UUT Channel 1 as follows: 1.009 DISP [32] 5520A NORMAL HI to A +INPUT TERMINAL 1.009 DISP [32] 5520A NORMAL LO to b -INPUT TERMINAL 1.010 JMP 1.012 # Dot type inputs 1.011 DISP Connect 5520A to the UUT Channel 1 as follows: 1.011 DISP [32] 5520A NORMAL HI to A +INPUT TERMINAL 1.011 DISP [32] 5520A NORMAL LO to B -INPUT TERMINAL 1.012 DISP Set the FUNCTION to DCV. 1.012 DISP Set RANGE to 20mV. 1.012 DISP Set CHANNEL # to 1. 1.013 5520 0.00mV 0.03U 2W 2.001 5520 20.00mV 0.2% 0.03U 2W 3.001 DISP Select the 60mV Range. 3.002 5520 0.00mV 0.02U 2W 4.001 5520 60.00mV 0.2% 0.02U 2W 5.001 DISP Set RANGE to 200mV 5.002 5520 0.0mV 0.2U 2W 6.001 5520 200.0mV 0.2% 0.2U 2W 7.001 DISP Set RANGE to 2V 7.002 5520 0.000V 0.1% 0.002U 2W 8.001 5520 2.000V 0.1% 0.002U 2W 9.001 DISP Set RANGE to 6V 9.002 5520 0.000V 0.002U 2W 10.001 5520 6.000V 0.3% 0.002U 2W 11.001 DISP Set RANGE to 20V. 11.002 5520 0.00V 0.02U 2W 12.001 5520 20.00V 0.3% 0.02U 2W 13.001 HEAD {RECORDING} 13.002 DISP Set RANGE to 2V. 13.003 DISP The following tests verify the recording PEN reflects 13.003 DISP the applied voltages. Check using a glass scale or 13.003 DISP ruler capable of .3mm measurements. 13.004 DISP Channel 1 ZONE must be set to 0~100mm 13.004 DISP by performing the following: 13.004 DISP 13.004 DISP Press the ESC key for 3 seconds. 13.004 DISP Press the UP/DOWN arrows until SET=AUX 13.004 DISP Press the ENTER key. Select MODE=ZONE,press ENTER. 13.004 DISP Set L = 0mm, R = 100mm. Press and hold the ESC key to 13.004 DISP return to normal operation. 13.005 DISP Press the RCD key to start the chart recorder 13.006 5520 -2.0000V S 2W 13.007 EVAL Is the {pen movement 0mm +/- .3mm}? 14.001 5520 -1.0000V S 2W 14.002 EVAL Is the {pen movement 25mm +/- .3mm}? 15.001 5520 0.0000V S 2W 15.002 EVAL Is the {pen movement 50mm +/- .3mm}? 16.001 5520 1.0000V S 2W 16.002 EVAL Is the {pen movement 75mm +/- .3mm}? 17.001 5520 2.0000V S 2W 17.002 EVAL Is the {pen movement 100mm +/- .3mm}? 18.001 DISP Press the RCD key to stop the recorder. 18.002 HEAD {REFERENCE JUNCTION COMPENSATION} 18.003 DISP Short circuit the High and Low terminals on the UUT 18.003 DISP using standard copper wire. 18.004 DISP Set the RANGE to the thermocouple type you wish to use 18.004 DISP (any TC is acceptable). 18.005 OPBR Is the selected thermocouple type R,S,B or W? 18.006 JMPF 19.002 18.007 EVAL Does the UUT indicate {AMBIENT temperature +/- 1degC}? 19.001 JMP 20.001 19.002 EVAL Does the UUT indicate {AMBIENT temperature +/-.5degC}? 20.001 HEAD {RTD} 20.002 DISP Connect the 5520A to the UUT Channel 1 as follows: 20.002 DISP [32] 5520A NORMAL HI to A INPUT TERMINAL 20.002 DISP [32] 5520A NORMAL LO to B INPUT TERMINAL 20.002 DISP [32] 5520A NORMAL LO to b INPUT TERMINAL 20.002 DISP All leads should be equal length and of 20.002 DISP the same type wire and connectors 20.003 DISP Set the FUNCTION to RTD. 20.003 DISP Set UP RJC to EXT 20.003 DISP Set reference value to 0uV 20.004 5520 -200.0degC 0.15% 0.3U R1 2W 21.001 5520 0.0degC 0.15% 0.3U R1 2W 22.001 5520 600.0degC 0.15% 0.3U R1 2W 23.001 HEAD {CHART SPEED} 23.002 DISP Set CHART SPEED to 1000mm/h. 23.002 DISP 23.002 DISP Start feeding chart by pressing RCD key. 23.002 DISP After one hour, stop feeding chart by 23.002 DISP pressing RCD key. 23.003 EVAL Does {chart measure between 999mm and 1001mm?} 24.001 END